Dosimetric Verification of the ADAC Pinnacle3 Pencil Beam Algorithm For Clinical Electrons In Presence of Cerrobend Blocking
by Chan, Philip, M.S., The University of Toledo, 2007, 75; 10630856
Abstract (Summary)

The purpose of this study was to commission planning electrons in the ADAC Pinnacle3treatment planning system and to verify its accuracy with clinical Cerrobend blocked fields. Pinnacle utilizes the Hogstrom electron pencil beam algorithm to perform its electron dose calculation. Complete set of measurement beam data in water phantom was first acquired and verified for consistency. The data was then entered into the planning system to develop a measurement and parameter-based model per beam. After the system was commissioned for clinical use, additional verification measurements were performed with unique field blocking. Comparisons between measured and computed dose from Pinnacle demonstrated the calculation was accurate and within clinical tolerance for all field blocking at standard treatment distance. For extended distance, relative depth dose prediction was poor only in the case with very high percentage field blocking. Relative cross-beam dose results were also poorer but were still acceptable for clinical use.

Indexing (document details)
Advisor:
Commitee: Dennis, Michael, Feldmeier, John, Parsai, E. Ishmael
School: The University of Toledo
Department: Biomedical Sciences
School Location: United States -- Ohio
Source: MAI 56/05M(E), Masters Abstracts International
Source Type: DISSERTATION
Subjects: Nuclear physics
Keywords: Adac pinnacle, Commissioning, Electron therapy, Hogstrom algorithm
Publication Number: 10630856
ISBN: 9780355011876
Copyright © 2019 ProQuest LLC. All rights reserved. Terms and Conditions Privacy Policy Cookie Policy
ProQuest